CHAPTER THREE
3-2
snugly with the target window opening centered over
the white ceramic disk.
2. Lower the instrument to the shoe; hold steady until
the user dialog indicates the calibration is completed.
NOTE: If an Optics Change? message appears during the white
calibration, select the No option and re-measure calibration
reference. If the instrument optics was changed, a full calibration
must be performed. Refer to the Full Calibration procedure that
follows.
3. Store the calibration reference in a dry, dust free area,
away from direct exposure to light.
CALIBRATION
White
Measure white
patch on
calibration
reference
<Measure White>
Calibration Reference
White Patch