INSTRUMENT FUNCTIONS
5-13
3. Press the Enter key # to alternate between Seq and
Match.
4. Center the target window over the first reference (or
replacement reference if a match), and lower the
instrument to take a reading.
5. Continue with additional reference measurements.
To manually edit reference values:
1. If not selected, press the Tab Up key $ or Tab Down
key @ to highlight Reference. Press the Enter key #
to access the reference menu.
NOTE: Reference will not appear in the measurement list unless
COLOR–REF01 is selected as the active density mode. Refer to Color
Measurement Mode earlier in this section.
2. Use the Tab keys $@ to move the highlight to the
desired reference location.
3. Press the Enter key # to move the highlight to the data
side of the screen. <
Enter Ref> appears in the user
dialog.
NOTE: Highlighting Clear Ref and pressing Enter key # can
quickly clear current reference data.
4. Press the Tab Up key $ or Tab Down key @ to
highlight desired attribute. Press and Enter key # to
access
Reference menu.
5. Use the Tab keys $@ to choose the desired digit
(arrows above and below designated selection). Press
the Enter key # to access the editor.
NOTE: Positioning the arrows above and below + or and pressing
the Enter key # toggles between the two symbols.
6. Use the Tab keys $@ to highlight the desired number
and press the Enter key # to exit the editor.
7. When editing is completed, use the Tab keys $@ to
highlight Save & Exit and press Enter key #.
Measuring a Color Sample
So far, you have performed the procedures to select the color
mode, options and illuminant, and to measure references.
Seitenansicht 66