CHAPTER SIX
6-4
Aperture and Polarization Kit Installation
(Excludes Micro-Spot instrument)
The 500 series instrument was designed to allow you to quickly
change the aperture and target window. X-Rite provides three
aperture kits especially designed for the 500 series instrument.
Available kits: 2mm, 3.4mm (standard), and 6mm.
To install the aperture and polarization kit:
1. Rotate shoe stop and open shoe as explained in the
battery replacement procedure (see previous page).
2. With instrument resting on the top cover, rotate
optics cap counter-clockwise until the triangle on the
cap meets the diamond on the bottom housing.
3. Carefully remove the optics cap by lifting upwards
and set aside.
4. Align aperture extraction tool over existing aperture as
shown. (Note alignment mark orientation—align with
side screw.)
Diamond Mark
Triangle Mark
Optics Cap
A
lignment Mark
Side Scre
w
Aperture
A
perture Tool